GG-S-1343A
November 26, 1975
SUPERSEDING
Int. Fed. Spec. GG-S-001343
December 15, 1969 and
Fed. Spec. GG-S-751a (In part)
November 9, 1940
FEDERAL SPECIFICATION
STERILIZER, SURGICAL INSTRUMENT AND SUPPLY
MECHANICAL AIR REMOVAL, NONPORTABLE
(HEAT AND MOISTURE STABLE)
This specification was approved by the Commissioner, Federal
Supply Service, General Services Administration, for use of all
Federal Agencies.
1. SCOPE AND CLASSIFICATION
1.1 Scope. This specification covers a sterilizer for processing heating and
moisture-stable materials by the mechanical-air-removal principle, using saturated steam as
the sterlizing agent.
1.2 Classification. The sterilizer shall be of the type whereby air is evacuated from
the chamber before and after the sterlizing exposure period by mechanical means. The
sterilizer shall be of the following size and style as specified (see 6.2).
1.2.1 Size. The minimum clearance dimensions of the chamber size of the inner shell
excluding space required for baffle plate (see 3.6.1.1) shall be as shown in table I.
TABLE I. Inner shell dimensions (inches)
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Size Width Height Length
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
3 24 36 36
4 24 36 48
5 24 36 60
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
1.2.2. Style.
Style A - Single door, cabinet enclosed (free standing).
Style B - Single door, for recessing through one wall partition.
Style C - Double door, (for goods pass-through); for recessing through one wall
partition (sizes 4 and 5).
Downloaded from http://www.everyspec.com